• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH Back to Search Results
Model Number 407200
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Embolus (1830)
Event Date 10/20/2017
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported clot formation could not be conclusively determined.Per the ifu, thromboembolism is a known risk during the use of this device.
 
Event Description
Related manufacturing ref: 3005334138-2017-00221.During a pulmonary vein isolation procedure for persistent atrial fibrillation, a clot formed on the tip of the introducer.Prior to transseptal puncture, a transesophageal echo was performed and revealed no clots.When the introducer was placed in the correct position for transseptal puncture, a transesophageal echo revealed a clot around the distal tip of the introducer in the right atrium.The procedure was cancelled and medication was administered to remove the clot.The patient is in stable condition.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7025405
MDR Text Key91832427
Report Number3008452825-2017-00274
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model Number407200
Device Catalogue Number407200
Device Lot Number6087243
Other Device ID Number05414734205092
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/20/2017
Initial Date FDA Received11/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SWARTZ¿ BRAIDED INTRODUCER
Patient Outcome(s) Other;
-
-