Model Number SC-8352-70 |
Device Problems
Fracture (1260); High impedance (1291); Material Separation (1562); Component Missing (2306)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/25/2017 |
Event Type
Injury
|
Event Description
|
A report was received that the patients lead was broken.It was mentioned that the lead came out the epidural space and one of the contacts was having high impedance.The patient underwent a lead replacement procedure.No device malfunction was suspected.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Sc-8352-70 (b)(4) device evaluation indicated that the complaint was confirmed.Visual inspection revealed that electrode #16 was dislodged and it was missing.Cable #16 was exposed.Additional visual inspection revealed the silicone where the paddle end met the four pigtails was damaged/torn.The damage was caused by excessive tensile force being exerted against the silicone material.Review of the device history record revealed no anomalies.Sc-4316 (ln 19964238) device evaluation indicated that both clik anchors passed all the required tests and revealed no anomalies.
|
|
Event Description
|
A report was received that the patients lead was broken.It was mentioned that the lead came out the epidural space and one of the contacts was having high impedance.The patient underwent a lead replacement procedure.No device malfunction was suspected.The patient was doing well postoperatively.
|
|
Manufacturer Narrative
|
Additional information was received that everything was removed from the patients body.
|
|
Event Description
|
A report was received that the patients lead was broken.It was mentioned that the lead came out the epidural space and one of the contacts was having high impedance.The patient underwent a lead replacement procedure.No device malfunction was suspected.The patient was doing well postoperatively.
|
|
Search Alerts/Recalls
|