• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY POWERLINE ULTRASONIC INSERT - 1000; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL 30K FOCUSED SPRAY POWERLINE ULTRASONIC INSERT - 1000; SCALER, ULTRASONIC Back to Search Results
Catalog Number 80799
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Unspecified Infection (1930); Burn, Thermal (2530)
Event Date 10/17/2017
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a 30k fsi-pwr-1000 insert, the insert was overheating and burned a patient's mucosa.Stomatitis was expected when burning occurred so the patient received medication.Stomatitis developed after a few days.
 
Manufacturer Narrative
The device was evaluated and found to be within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FOCUSED SPRAY POWERLINE ULTRASONIC INSERT - 1000
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
1301 smile way
york PA 17404
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7029622
MDR Text Key91959976
Report Number2424472-2017-00216
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052334
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80799
Device Lot Number17191
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/17/2017
Initial Date FDA Received11/14/2017
Supplement Dates Manufacturer Received11/22/2017
Supplement Dates FDA Received12/18/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-