Catalog Number FNC1168
Device Problem Cut In Material (2454)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was received for evaluation.Visual inspection revealed a cut in the tubing at the connection to the upper portion of the y-junction.The tubing was nearly cut off of the junction.The cause of the condition was determined to be a manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation of a returned blood/solution/anesthesia set, a cut was found in the tubing at the connection to the upper portion of the y-junction.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
singapore
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7029738
MDR Text Key92905303
Report Number1416980-2017-09052
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberFNC1168
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2017
Initial Date Manufacturer Received 10/18/2017
Initial Date FDA Received11/14/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N