An event regarding crack/fracture involving a scorpio insert was reported.The event was confirmed.A material analysis of the returned device conducted on (b)(6) 2017 concluded "post fracture abrasion obscured the fracture surface; no fracture surface analysis was possible.Abrasion damage was observed on the posterior portion of the post, consistent with cyclic contact with the femoral component.Burnishing, scratching, and third-body indentations were observed on the articulating surface, consistent with commonly-identified damage modes in uhmwpe inserts.No material or manufacturing defects were observed on the device features examined." no patient medical records were available for review.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.There has been no other event for the lot referenced.A material analysis of the returned device confirmed the event as abrasion damage consistent with cyclic contact with the femoral component was observed on the posterior portion of the post.However, post fracture abrasion obscured the fracture surface and hence no fracture surface analysis was possible.Burnishing, scratching, and third-body indentations consistent with commonly-identified damage modes in uhmwpe inserts were also observed on the articulating surface.No material or manufacturing defects were observed on the device features examined.
|