Catalog Number 11003-59
Device Problems Device Damaged by Another Device (2915); Material Deformation (2976)
Patient Problems Occlusion (1984); Thrombosis (2100)
Event Date 10/26/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The right iliac was successfully stented with a 7x100mm non-abbott stent being placed over the omnilik elite from the aortoiliac bifurcation to the external iliac artery.The situation worsened as the right side was opened, but there was no flow due to manipulations with the 7f introducers and the 8f sheath.The case was completed with manual compression to the right common femoral artery due to the calcification, which caused a cuff-miss to occur when a proglide was used.After a day the patient had to undergo surgery as the stents were again occluded.The occlusion was a mixture of thrombosis and mesh.The patient is doing well after undergoing surgery.No additional information was provided.Concomitant medical products: dilatation catheter: 6x60mm armada 35; guide wire: storque 35 cordis; sheath: cordis 7f.The device was not returned for evaluation.The reported patient effects of occlusion and thrombosis are a known observed and potential patient effects as listed in the omnilink elite instructions for use.A review of the lot history record and complaint history of the reported device could not be conducted because the lot number was not provided.The investigation determined the reported difficulties, patient effects and treatment appear to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.The 6x60mm armada 35 and proglide devices are being filed under separate medwatch reports.
 
Event Description
It was reported that the procedure was to treat a 100% stenosed and heavily calcified restenosed lesion in the external iliac artery.Following pre-dilatation, a 7x59mm omnilink elite stent was implanted.The stent had 15% residual restenosis; therefore, a 6x60mm armada 35 balloon catheter was advanced and inflated past the rated burst pressure to fully appose the stent.The balloon ruptured circumferentially.During an attempt to remove the ruptured balloon parts of the balloon became stuck with the struts of the implanted stent and caused the balloon catheter to separate.The proximal part of the balloon catheter was removed.To remove the entrapped distal portion of the balloon, a non-abbott wire was advanced on the left side and an 8f introducer was crossed over to the right common iliac artery.The non-abbott wire was pulled through from left to right; however, the tip of the wire got stuck in the previous implanted stent resulting in stent struts becoming flared.An unknown 5mm dilatation catheter was advanced via the left side and while performing successive dilatation of the balloon, the separated distal part of the armada 35 balloon catheter was removed from the stent struts.The separated part of the balloon was then fixed using a 70cm-6f-sheath from the right side and an 8f sheath from the left side.Via a long maneuver with push and pull, the separated part of the balloon catheter was removed through a 7f into the 8f sheath out of the body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Type of Device
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7039741
MDR Text Key92289645
Report Number2024168-2017-09053
Device Sequence Number1
Product Code NIO
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P110043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 11/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11003-59
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/30/2017
Initial Date FDA Received11/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-