MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Only year (2017) is known.Device evaluation: the analysis results found that the ats45 device was received with no apparent damaged and with two reloads present not loaded on the device.Cartridge b, (b)(4), partially fired 1/10, lockout normal.Cartridge c, (b)(4), partially fired 1/10, lockout normal.This indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.The returned device was tested for functionality with a test reload and it fired, cut, and all the staples formed as intended.The device fired without any difficulties, the staple line was complete and the staples were noted to have the proper b-formed shape.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during an unknown procedure the stapler presented a defect at the moment of firing the loads - white and blue.It occurred twice with two charges.It was necessary to replace the loads.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS FLEX ARTICNG LNR CUTR 45MM
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7040182
MDR Text Key93060455
Report Number3005075853-2017-06164
Device Sequence Number1
Product Code GDW
Combination Product (y/n)N
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/04/2021
Device Catalogue NumberATS45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2017
Initial Date Manufacturer Received 10/25/2017
Initial Date FDA Received11/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-