(b)(4).Only year (2017) is known.Device evaluation: the analysis results found that the ats45 device was received with no apparent damaged and with two reloads present not loaded on the device.Cartridge b, (b)(4), partially fired 1/10, lockout normal.Cartridge c, (b)(4), partially fired 1/10, lockout normal.This indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.The returned device was tested for functionality with a test reload and it fired, cut, and all the staples formed as intended.The device fired without any difficulties, the staple line was complete and the staples were noted to have the proper b-formed shape.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|