It was reported that the customer tightened the adapter of the contamination shield excessively after the operation at icu, and therefore ra (right atrium) pressure was increased more than expected.At that time proximal injectate lumen was connected to the blood pressure kit and ra pressure was monitored.The lumen was kinked due to over-tightening, and therefore they suspected it caused the occlusion of the lumen.The value of ra pressure became normal when the adapter of the contamination shield was loosened.The pulmonary artery (pa) pressure value had been normal.Occurrence date is unknown.There were no patient complications reported.Additional information was reported that the indicated cvp value was over 15mmhg, while the expected value was unknown.The value and waveform matched, the patient was not treated based on the incorrect value.The error message was not observed.The value was not affected by the patient condition.Patient demographic information requested but unavailable.
|
One catheter with detached monoject 1.5 cc limited volume syringe and two three-way stopcocks were returned for evaluation.A non-edwards contamination shield was located on the catheter between 41.5 cm and 97 cm proximal from the catheter tip.Clotted blood was observed from the contamination shield.Without the return of the pressure monitoring unit, the customer report of pressure measurement issue could not be confirmed.All through lumens were patent without any leakage or occlusion.No resistance was felt when air or water was injected to the through lumens when the contamination shield was loosened or tightened.A lab 0.018" angle tip guidewire (as recommended in the ifu) was passed hub to tip and tip to hub without resistance when the contamination shield was loosened or tightened.After removing the contamination shield, three indentations were found at 42.5cm, 94.5 cm and 95cm.It was able to insert the catheter without resistance into a lab 8.5fr and 9fr introflex introducer which is the recommended size introducer per the ifu.No visible damage or inconsistency to the balloon, or returned syringe was observed.The balloon inflated clear, concentric and remained inflated for more than 5 minutes without leakage.Balloon inflation testing was performed using returned syringe with 1.5 cc air and a calibrated stopwatch.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.Customer report of pressure measurement issue could not be confirmed during the analysis, as the device responded appropriately during functional testing.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.With any hemodynamic monitoring, pressure readings can change quickly and dramatically.Pressure readings should correlate with the patient¿s clinical manifestations.In this case it is unknown whether any user or procedural factors may have contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
|