Model Number 8011-0501-01 |
Device Problem
Failure to Discharge (1169)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during biomed testing, the associated defibrillator failed to discharge using these internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The malfunction was duplicated and attributed to a faulty switch on the internal handles.The internal handles were scrapped.Analysis of reports of this type has not identified an increase in trend.
|
|