Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Review of device history records show the lot released with no recorded anomaly or deviation.Current information is insufficient to permit a valid conclusion about the cause of this event; a follow-up report will be sent upon completion of the device evaluation.
 
Event Description
It was reported the wire cutter was identified to be chipped.The office is uncertain if the chip happened during surgery.The chip was noticed during surgery; they looked to see if they had another wire cutter but they did not, therefore the surgeon continued to use the chipped wire cutter to complete the procedure.There was no surgical delay in excess of 30 minutes, no injury, and no foreign body in the patient.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.The instrument was visually evaluated and found to be fractured at the working end.The rest of the instrument is in good overall cosmetic condition with normal scratches and wear from use; no discoloration was observed.There were no indications of manufacturing defects.Investigation results concluded that the reported event was due to excessive force either being applied during use or from being dropped or mishandled.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMITH WIRE CUTTER WITH NOTCH
Type of Device
WIRE CUTTER
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
michelle cole
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key7046644
MDR Text Key93247876
Report Number0001032347-2017-00816
Device Sequence Number1
Product Code HXZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PN/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Dentist
Device Model NumberN/A
Device Catalogue Number51-0929
Device Lot Number102516J16
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2017
Initial Date FDA Received11/20/2017
Supplement Dates Manufacturer Received04/10/2018
Supplement Dates FDA Received04/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-