• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ5 HI OFF; HIP FEMORAL STEM/SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT POR TAPER SZ5 HI OFF; HIP FEMORAL STEM/SLEEVE Back to Search Results
Catalog Number 157011110
Device Problems Corroded (1131); Appropriate Term/Code Not Available (3191)
Patient Problems Host-Tissue Reaction (1297); Foreign Body Reaction (1868); Pain (1994); Skin Irritation (2076); Swelling (2091); Tissue Damage (2104)
Event Date 05/06/2015
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Litigation alleges the patient suffers from pain, discomfort and toxic cobalt-chromium metal ions and particles to be released into the blood, tissue and bone.In addition to what was previously alleged, pfs alleges swelling, drainage, large hematoma,yellow skin color, grinding and a cane for walking.After review of medical records for the mdr reportability, patient was revised to address failed right total hip arthroplasty due to metallosis, severe pain, dysfunction and failed recall stem implant.Revision notes reported of increased pain, dysfunction, peritrochanteric soft tissue mass and elevated metal lie.It was also indicated that there was an extensive scar tissue and changes in the soft tissues in the peritrochanteric tissues.It was also stated that there was a large grey metallosis, metal corrosion, extra bone in anterior-inferior recess and malpositioned liner.It was noted within the revision note that their was evidence of cup loosening.
 
Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ5 HI OFF
Type of Device
HIP FEMORAL STEM/SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581-0988
5743725905
MDR Report Key7053985
MDR Text Key92793242
Report Number1818910-2017-29623
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295060062
UDI-Public10603295060062
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Attorney
Type of Report Initial,Followup
Report Date 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/01/2018
Device Catalogue Number157011110
Device Lot NumberC3JBG1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/26/2017
Initial Date FDA Received11/22/2017
Supplement Dates Manufacturer Received09/10/2018
Supplement Dates FDA Received09/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient Weight86
-
-