LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-88-00
Device Problem Failure to Pump (1502)
Patient Problem No Patient Involvement (2645)
Event Date 10/23/2017
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to two defective boards.The technician replaced the motor control and motor end stage board to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The defective boards were returned to livanova (b)(4) for further investigation.During the evaluation of the boards the reported failure could be confirmed.It was possible to trace the failure back to a software issue on the motor control board.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
Livanova (b)(4) received a report that a s5 mast roller pump did not turn on during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7056652
MDR Text Key93170308
Report Number9611109-2017-00918
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K062396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 11/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-88-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/23/2017
Initial Date FDA Received11/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-