• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH Back to Search Results
Model Number G407208
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 11/02/2017
Event Type  Injury  
Event Description
Related manufacturing reference numbers: 9680001-2017-00091, 3005334138-2017-00236, 3005334138-2017-00237.During an atrial fibrillation ablation procedure, a pericardial effusion occurred.Following ablation of the cavotricuspid isthmus at the end of the procedure, the catheter was advanced to the left atrium and the patient became hypotensive.An echocardiogram revealed an effusion, for which a pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott device.
 
Manufacturer Narrative
A pericardial effusion was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7063174
MDR Text Key93069648
Report Number3008452825-2017-00308
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Model NumberG407208
Device Lot Number6081482
Other Device ID Number05414734205139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2017
Initial Date FDA Received11/28/2017
Supplement Dates Manufacturer Received11/28/2017
Supplement Dates FDA Received11/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ADVISOR FL CATHETER; SWARTZ BRAIDED INTRODUCER; TACTICATH ABLATION CATHETER
Patient Outcome(s) Required Intervention;
Patient Age80 YR
Patient Weight74
-
-