Model Number 500AVHCT
Device Problems No Display/Image (1183); Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/03/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) found that the auxiliary board had an overheated capacitor.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the monitor would not power on.A blood gas analyzer was used for monitoring.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7064134
MDR Text Key93605403
Report Number1828100-2017-00534
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number500AVHCT
Device Catalogue Number500AVHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/07/2017
Initial Date FDA Received11/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-