ZIMMER BIOMET, INC. FEMORAL HEAD PROVISIONAL 12/14 NECK TAPER; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problems Disassembly (1168); Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Concomitant: 00785001300, versys advocate femoral stem, 62955177.(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2017 - 00724.
 
Event Description
It was reported that when the surgeon was trying to pull off the head provisional from the stem the stem came out of the femur.The surgery was finished with a backup stem.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.A versys advocate femoral stem and a versys femoral head provisional were returned for evaluation.As returned, the stem is in excellent condition and shows no signs of damage to the taper.The femoral head provisional exhibits damage to the outer sphere and conical taper.Damage is too severe for dimensional analysis.Review of the device history record identified no deviations or anomalies would contribute to reported event.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD PROVISIONAL 12/14 NECK TAPER
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7064888
MDR Text Key93561267
Report Number0001822565-2017-08094
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number00789502602
Device Lot Number60325007
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/02/2017
Initial Date FDA Received11/28/2017
Supplement Dates Manufacturer Received10/30/2018
Supplement Dates FDA Received11/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-