• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-6201A
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2017
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reported that the central nurse's station (cns) went down after a power outage and came back up to a blue screen.Switching the hard disk drives (hdd) around allowed the primary drive to rebuild, which resolved the issue.The device was in use on patients at the time, however no patient harm was reported.
 
Event Description
The biomedical engineer reported that the central nurse's station (cns) went down after a power outage and came back up to a blue screen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6201A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key7070994
MDR Text Key93493003
Report Number8030229-2017-00415
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921114131
UDI-Public4931921114131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6201A
Device Catalogue NumberPU-621RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/30/2017
Distributor Facility Aware Date11/03/2017
Device Age34 MO
Event Location Hospital
Date Report to Manufacturer11/30/2017
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received11/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-