NIHON KOHDEN CORPORATION CNS-6201A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-6201A
Device Problems No Audible Alarm (1019); Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/31/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the central nurse's station (cns) alarmed tachycardia, however no audible alarm sounded.While reviewing the log files, it was discovered that the silence alarm button had been pressed, which is the reason no alarm sounded for this event.The customer was advised that pressing the silence alarm button will silence all alarms for 2 minutes.No patient harm was reported as a result of this event.
 
Event Description
The customer reported that the central nurse's station (cns) alarmed tachycardia, however no audible alarm sounded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6201A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key7071004
MDR Text Key93493361
Report Number8030229-2017-00419
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921114131
UDI-Public4931921114131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6201A
Device Catalogue NumberPU-621RA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA11/30/2017
Distributor Facility Aware Date10/31/2017
Device Age19 MO
Event Location Hospital
Date Report to Manufacturer11/30/2017
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received11/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-