The returned device, intended for use in treatment, was returned for evaluation.There was a relationship found between the device and the reported incident.Visual inspection shows a broken/detached grasper at distal end.The grasper was not returned for further evaluation.During the functional evaluation both needles could be extracted by using a smart stitch device.The rod inside the pp-connector doesn't move when the clamp lever is pressed.The complaint was verified.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|