• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD INSTRUMENT SEDI-40; SEDIMENTION INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD INSTRUMENT SEDI-40; SEDIMENTION INSTRUMENT Back to Search Results
Catalog Number 361546
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2017
Event Type  malfunction  
Manufacturer Narrative
Medical device lot # unknown.Medical device expiration date: unknown.(b)(6).Device manufacture date: unknown.Results: a sample was returned for evaluation, the instrument was tested and was found to be incorrectly calibrated ¿ this was corrected and a new fan installed as the existing one was noisy.Conclusion: root cause is indeterminate.
 
Event Description
It was reported that bd instrument sedi-40 had llo messages and inadequate filling upon use.No medical intervention or injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSTRUMENT SEDI-40
Type of Device
SEDIMENTION INSTRUMENT
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7081305
MDR Text Key94409365
Report Number2243072-2017-00296
Device Sequence Number1
Product Code GHC
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number361546
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/05/2017
Initial Date FDA Received12/04/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-