• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL ENHANCED PLATELET, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL ENHANCED PLATELET, PLASMA, RBC SET Back to Search Results
Catalog Number 80440
Device Problems Device Operates Differently Than Expected (2913); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Patient Involvement (2645)
Event Date 11/09/2017
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process.A followup report will be provided.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Wbc count is not available at this time.The platelet collection is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information.The customer did not provide the lot number pertaining to this event,therefore a device history record (dhr) search could not be conducted for this specific incident.All lots must meet acceptance criteria before release.There was no run data file (rdf) available for analysis, therefore, root cause remains undetermined at this time.Possible causes for the elevated wbc count include the following: rbc spill over.Centrifuge stopped.Rbc detector calibration error.Possible air block.Donor physiology.Sampling, calculation, or other process error.
 
Event Description
The customer did not respond to attempts to obtain information for the investigation, such asprocedural details, wbc count and lot information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL ENHANCED PLATELET, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
steve kern
10810 w. collins ave
lakewood, CO 80215
3032392246
MDR Report Key7083062
MDR Text Key95078299
Report Number1722028-2017-00464
Device Sequence Number1
Product Code GKT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK170059
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80440
Other Device ID Number05020583804401
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2017
Initial Date FDA Received12/05/2017
Supplement Dates Manufacturer Received12/07/2017
Supplement Dates FDA Received12/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-