Returned product consisted of a ffr comet pressure wire.The tip, device shaft, sensor port and the coefficient values was examined for damage or any irregularities.The shaft showed a kink.The kink was approximately 61.5cm from the tip.There was also some slight peeling of the coating on the kinked area.The tip showed no damage.The sensor port was clear of any material.Inspection of the remainder of the device revealed no other damage or irregularities.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire.The coefficient was confirmed to be in specification.The wire functioned as designed.Because there was no evidence of any product quality deficiencies, it was considered likely that the kink and the peeling of the coating were attributable to handling of the device.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
Reportable based on device analysis completed on 10nov2017.It was reported that a comet wire was not used due to ffr link malfunction.During the preparation outside the patient, a comet pressure guidewire was opened to perform fractionated flow reserve (ffr) measurement.Then it was noticed that the ffr link would not switch on.Different plug points and power cables were tried but would not work.The procedure was not completed due to this event.No patient complications were reported and the patient¿s status was stable.However, the returned device revealed coating peeled off.
|