Catalog Number 9734679
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.Part not returned for evaluation.
 
Event Description
A medtronic representative reported that during spinal fusion procedure, the tip of the lumbar probe bent.It was reported that the patient had hard bone and when the instrument was pulled from vertebral body.The tip appeared twisted.Issue occurred when in patient anatomy.The procedure was completed with the use of thoracic probe.There was no delay to procedure.No impact on patient outcome.
 
Manufacturer Narrative
The suspect probe was returned to the manufacturer for evaluation.Testing found that the probe was bent at the tip.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAV LUMBAR PROBE TIP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7087738
MDR Text Key93886717
Report Number1723170-2017-04970
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734679
Device Lot Number170406
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/08/2017
Initial Date FDA Received12/06/2017
Supplement Dates Manufacturer Received04/06/2018
Supplement Dates FDA Received04/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age57 YR
Patient Weight93
-
-