The universal cut block was returned and evaluated.A visual inspection found that the device had sharp burs around both sides of the cutting guide.Some abrasions were also noticed throughout the device.Additionally, a dimensional inspection was attempted; however the damage and deformation at several features of the device would not allow for accurate measurement.Finally, clinical investigation concluded the following: without sufficient clinically relevant supporting documentation, a thorough medical assessment and/or causal relationship cannot be provided.The patient impact beyond possible instability from the reported poorly maintained flexion/extension gap and the 45 minute surgical extension cannot be concluded.No further medical assessment is warranted at this time.We recommend that all reusable instruments be routinely inspected for wear and damage so they can be replaced as necessary.Reviewing the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.
|