|
Model Number RBY2C0320-A |
Device Problems
Kinked (1339); Device Damaged Prior to Use (2284); Device Handling Problem (3265)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the pet lock was intact on the proximal end of the pusher assembly for the first ruby coil evaluated.The pusher assembly was kinked in multiple locations.The embolization coil was intact with its pusher assembly.Conclusions: evaluation of the first returned device revealed that the ruby coil¿s pusher assembly was kinked in multiple locations.This damage was likely accidental during advancement of the ruby coil into the lantern on the back table as mentioned in the complaint.The kinks in the pusher assembly prevented the returned ruby coil from being functionally tested through a demonstration microcatheter.Evaluation of the returned device revealed that the podj¿s pet lock was broken and the pull wire was retracted out of the ddt.This type of damage typically occurs due to improper handling during use.If pet lock is broken and the pull wire is retracted out of the ddt during manipulation of the device, by design, the embolization coil will detach from its pusher assembly.The pull wire being retracted out of the ddt likely caused the embolization coil to detach from its pusher assembly.Evaluation of the second returned device revealed that the ruby coil was functional.The ruby coil was advanced through a demonstration microcatheter and out of the distal tip without an issue.Evaluation of the third returned device revealed that the ruby coil¿s pusher assembly was kinked and the introducer sheath had coagulated blood inside.The kink in the pusher assembly was likely accidental during advancement of the ruby coil into the lantern as mentioned in the complaint.The coagulated blood prevented the ruby coil from being advanced through the demonstration microcatheter during functional testing.Evaluation of the returned lantern revealed that the microcatheter distal shaft was repeatedly kinked.This type of damage likely occurred due to forceful manipulation during placement of the ruby coils.These kinks likely contributed to the ruby coil unintentionally detaching within the microcatheter.A mandrel could not be completely advanced through the returned lantern during functional testing.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2017-02119, 3005168196-2017-02120, 3005168196-2017-02122.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure using pod packing coils (podjs) and a lantern delivery microcatheter (lantern).During preparation for the procedure, the hospital technician accidentally kinked two ruby coils while advancing them into the lantern on the back table.A third ruby coil also became kinked during preparation on the back table.The damage to these ruby coils occurred prior to use, and therefore they were not used in the procedure.During the procedure, the physician successfully placed ruby coils using the same lantern.The physician then experienced difficulty placing a podj within the target location and, consequently, the podj unintentionally detached as the physician was manipulating it.The physician therefore removed the lantern with the podj inside, and the procedure was completed using a new lantern and additional ruby coils.There was no report of an adverse effect to the patient.
|
|
Manufacturer Narrative
|
Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #01 mfr report:3005168196-2017-02121.This report is associated with mfr report numbers: 3005168196-2017-02119, 3005168196-2017-02120, 3005168196-2017-02122.
|
|
Search Alerts/Recalls
|
|
|