• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problem Pacing Problem (1439)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Product event summary: analysis was unable to confirm the customer comment of programmer switching over to its emergency mode when device detection was "on", however the emergency button connections were checked and reseated.It was noted that the programmer failed incoming test 1012 (common mode/wrist electrode test, though this was not related to the customer complaint.The link electronic module board was replaced to resolve.The power cord bay was also noted to be broken and was also replaced.The circuit boards and mechanical parts were inspected, the hard drive reconfigured, the software reloaded and updated and the device then passed all final functional and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during an implant procedure three times in a row when the new (implantable heart) device's detection was "on" the programmer automatically switched to its "emergency" mode.It was further reported that this occurred while using the analyzer.The programmer was changed out and returned for service as well as a requested test and calibration procedure.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7090283
MDR Text Key94492272
Report Number2182208-2017-01876
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2017
Initial Date Manufacturer Received 10/16/2017
Initial Date FDA Received12/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-