MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENRHYTHM DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENRHYTHM DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number P1501DR
Device Problems Battery Impedance Issue (2884); Battery Problem (2885); Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/23/2017
Event Type  Injury  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.
 
Event Description
It was reported that the implantable pulse generator (ipg) experienced an electrical reset, which removed a software update that provided additional battery measurements.Subsequently, the device did not trigger elective replacement indicator (eri).In addition, the battery impedance was noted to have started to increase recently, although the battery voltage was stable.Reprogramming was performed, and the device remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENRHYTHM DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7090288
MDR Text Key94070680
Report Number3004209178-2017-25013
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/14/2012
Device Model NumberP1501DR
Device Catalogue NumberP1501DR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2017
Initial Date FDA Received12/07/2017
Date Device Manufactured04/25/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1440-2010
Patient Sequence Number1
Treatment
(B)(4) LEAD, (B)(4) LEAD
Patient Outcome(s) Required Intervention;
Patient Age93 YR
-
-