• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO REVO MRI SURESCAN; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO REVO MRI SURESCAN; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE Back to Search Results
Model Number RVDR01
Device Problems Incorrect Measurement (1383); Inappropriate or Unexpected Reset (2959); Missing Test Results (3267)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/30/2017
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the implantable pulse generator (ipg) device displayed invalid battery data due to a suspected bit-flip.Device data confirmed that a power on reset (por) occurred.The patient will be brought in to the clinic to clear the reset observation.The device remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a power on reset occurred.Analysis of the device memory indicated the battery measurement was not available.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the implantable pulse generator (ipg) device displayed invalid battery data due to a suspected bit-flip.Device data confirmed that a power on reset (por) occurred.The patient will be brought in to the clinic to clear the reset observation.The device remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVO MRI SURESCAN
Type of Device
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7091255
MDR Text Key94056087
Report Number3004209178-2017-25135
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00613994614742
UDI-Public00613994614742
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/14/2014
Device Model NumberRVDR01
Device Catalogue NumberRVDR01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/31/2017
Initial Date FDA Received12/07/2017
Supplement Dates Manufacturer Received01/23/2018
Supplement Dates FDA Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5086MRI52 LEAD, 5086MRI58 LEAD
Patient Outcome(s) Required Intervention;
Patient Age73 YR
-
-