MEDTRONIC MEXICO BROCKENBROUGH NEEDLE; TROCAR Back to Search Results
Model Number EP003994S
Device Problems Break (1069); Use of Device Problem (1670)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/18/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during an radiofrequency procedure, during the transseptal procedure, when the needle was removed from the introducer, there was material on the tip of the needle.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the transseptal needle, ep003994s was not returned.While no product was returned, an analysis of an image showed evidence of skiving.It is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key7092854
MDR Text Key94666004
Report Number9612164-2017-01770
Device Sequence Number1
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public00613994663146
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/27/2019
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number213532707
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2017
Initial Date FDA Received12/07/2017
Supplement Dates Manufacturer Received02/10/2020
Supplement Dates FDA Received03/11/2020
Date Device Manufactured07/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-