Model Number 2090W |
Device Problems
Device Emits Odor (1425); Power Problem (3010)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/04/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis:the reports of inability to power on and foul odor were confirmed.The programmer was found to have a faulty power supply.The programmer was also found to have a loose ecg (electrocardiogram) connector on the circuit board, which was out of electrical specification.The programmer is no longer needed and will not be repaired or returned to use.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that, during a device interrogation, the programmer powered off.A foul, "hot" smell was noted at the time of the issue.The programmer was returned for service.No patient complications have been reported as a result of this event.
|
|