• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. SMARTSTITCH PP CONNECTOR; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SMARTSTITCH PP CONNECTOR; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number OM-8010
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2017
Event Type  malfunction  
Event Description
It was reported that the jaw broke off of two devices during use.The broken piece(s) were retrieved and the procedure was completed using a back-up device.No patient injury or other complications were reported.Complaint 1 of 2.See (b)(4) for 2nd device.
 
Manufacturer Narrative
The reported perfect passer connectors, intended for use in treatment, were returned for evaluation.A relationship between the devices and reported incident was established.From the information provided, jaws disengaged from the device.Visual inspection shows the connectors were received with the s-clamp unhooked from the s-hook.One of the connectors was returned without the s-clamp.Internal investigation indicates the failure cause for upper s-clamp coming loose is due to a dimensional discrepancy on the s-hook component.Changes to the component have been implemented to prevent this failure.The instruction for use (ifu) were reviewed and were found to outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSTITCH PP CONNECTOR
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
MDR Report Key7094656
MDR Text Key94874705
Report Number3006524618-2017-00401
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123268
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberOM-8010
Device Lot Number1194763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2017
Initial Date FDA Received12/07/2017
Supplement Dates Manufacturer Received12/22/2017
Supplement Dates FDA Received12/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-