• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNOVIS SURGICAL INNOVATIONS VASCULAR PROBES; DILATOR, VESSEL, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNOVIS SURGICAL INNOVATIONS VASCULAR PROBES; DILATOR, VESSEL, SURGICAL Back to Search Results
Catalog Number 7151520
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 08/07/2017
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was particulate matter in the inner pouch of a vascular probe.This was noticed before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection was performed and particulate matter (pm) was identified outside the wall of the inner pouch.Inspection under the microscope was performed and the pm was determined to fiber; however, the size of the pm was small enough to meet specification.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASCULAR PROBES
Type of Device
DILATOR, VESSEL, SURGICAL
Manufacturer (Section D)
SYNOVIS SURGICAL INNOVATIONS
saint paul MN
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7095156
MDR Text Key94224692
Report Number1416980-2017-09612
Device Sequence Number1
Product Code DWP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K130896
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number7151520
Device Lot NumberSP17F271237157
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/13/2017
Initial Date FDA Received12/07/2017
Supplement Dates Manufacturer Received01/15/2018
Supplement Dates FDA Received01/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-