• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION AUTOCLAVABLE INTERNAL HANDLES (MSERIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION AUTOCLAVABLE INTERNAL HANDLES (MSERIES) Back to Search Results
Model Number 8011-0501-01
Device Problem Failure to Discharge (1169)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that during incoming inspection, the associated device failed to discharge using these internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The paddles were returned to zoll medical corporation; the malfunction was duplicated and attributed to the paddles discharge switch.The paddles were scrapped and the customer received a replacement.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCLAVABLE INTERNAL HANDLES (MSERIES)
Type of Device
INTERNAL HANDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key7101109
MDR Text Key94284212
Report Number1220908-2017-03089
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
K963781
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8011-0501-01
Device Catalogue Number8011050101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/18/2017
Initial Date FDA Received12/08/2017
Supplement Dates Manufacturer Received11/18/2017
Supplement Dates FDA Received01/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-