• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523RNAS
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/11/2017
Event Type  malfunction  
Manufacturer Narrative
Pump was received with minor scratched lcd window, broken reservoir tube lip, missing reservoir tube o ring, cracked case at the reservoir tube window corners, cracked belt clip slot, missing end cap sticker, stained address/serial number label and broken battery tube threads.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.
 
Event Description
The customer was reported via phone call that, the chips of plastic coming off the top of the reservoir and battery compartments and the black o-ring is coming out of the reservoir compartment of the pump.The blood glucose was unknown at the time of the incident.Customer advised to replace and discontinue use of the pump and revert to a back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key7102314
MDR Text Key94378303
Report Number2032227-2017-71492
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169221550
UDI-Public(01)00643169221550
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523RNAS
Device Catalogue NumberMMT-523RNAS
Device Lot NumberA000290041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/22/2017
Initial Date FDA Received12/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age69 YR
Patient Weight148
-
-