MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-554WWL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-554WWL |
Device Problems
Break (1069); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/23/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states. however, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
The customer reported via phone call that their insulin pump was damaged.The patient¿s blood glucose level was unknown at the time of the incident.Reportedly, ring had come off, reservoir unable to click into place.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device is expected to be returned for analysis.
|
|
Manufacturer Narrative
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.
|
|