MAUDE Adverse Event Report: CYBERONICS, INC. LEAD MODEL 302

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2017
Event Type  malfunction  
Event Description
It was reported that this patient was having a generator replacement due to eos.During surgery the surgeon noticed that the lead was loose and not working due to observed high impedance.Therefore the physician also replaced the lead and the surgery became a full system replacement.The patient was said to have had a double mastectomy recently and the physician thought this may have been the reason for the issue.During the left mastectomy it was reported that the general surgeon had to move the generator during surgery.The device has not been received by the manufacturer to date.No other relevant information has been received to date.
 
Event Description
It was reported that the product was discarded in pathology at the hospital facility by mistake.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7104282
MDR Text Key94408559
Report Number1644487-2017-04970
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2008
Device Model Number302-20
Device Lot Number1403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 11/16/2017
Initial Date FDA Received12/11/2017
Supplement Dates Manufacturer Received12/15/2017
Supplement Dates FDA Received01/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
-
-