• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Low Battery (2584); Device Or Device Fragments Location Unknown (2590); Device Operates Differently Than Expected (2913)
Patient Problems Therapeutic Effects, Unexpected (2099); No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2016
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient who was implanted with an internal neurostimulator (ins) for urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor.It was reported that the caller stated that the patient¿s implant was not working.No further complications reported.No complications reported anticipated.No symptoms reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider regarding a patient who was implanted with a neurostimulator (ins) for gastrointestinal/pelvic floor and urinary dysfunction/sacral nerve stim.It was reported that the therapy was not working and the battery was dead.No new information.Patient reported again that patient's ins was dead and from at least last year.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7106441
MDR Text Key94839066
Report Number3004209178-2017-25791
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2017
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2017
Initial Date FDA Received12/11/2017
Supplement Dates Manufacturer Received12/04/2017
Supplement Dates FDA Received12/21/2017
Date Device Manufactured07/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-