Device Problems Signal Artifact/Noise (1036); Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported artifacts on images.There was no report of misrepresentation as a result of the ring artifacts.The philips field service engineer (fse) went to the site to evaluate the system.The fse visually inspected the system and found a crack in the compensator, which caused the artifacts.The fse replaced the a-plane collimator to resolve the issue.The fse confirmed there was no impact to a patient and no report of misrepresentation and/or mistreatment as a result of this reported issue.The system is in clinical use.This issue was determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRILLIANCE AIR 40/64/UCT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7110022
MDR Text Key94963553
Report Number1525965-2017-00126
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 07/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/04/2016
Initial Date FDA Received12/12/2017
Supplement Dates Manufacturer Received07/04/2016
Supplement Dates FDA Received11/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-