Model Number 728306
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.(b)(4).
 
Event Description
This complaint has been evaluated based on the information provided; there is no allegation of death or serious injury.The customer reported that they cannot hear the patient.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported that they could not hear the patient.The customer confirmed that there was no patient impact and no harm as a result of this event.The case was dispatched to a philips field service engineer (fse), who evaluated the ct system, and determined that the microphone in the breathing light assembly had failed.The fse replaced the breathing light assembly to resolve the reported issue.The system was handed back to the customer for clinical use after successful completion of the corrective maintenance.The system is operational and in clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
Manufacturer Contact
derrick massuri
595 miner rd
cleveland, OH 44143
4404833000
MDR Report Key7110182
MDR Text Key95074582
Report Number1525965-2017-00135
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/15/2017
Initial Date FDA Received12/12/2017
Supplement Dates Manufacturer Received11/15/2017
Supplement Dates FDA Received04/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-