For analysis at manufacturer site no log file was provided but it was stated by the biomed looking at the logs that no error entry was found.This fits to the statement of the customer that the device did not alarm or go into vent fail during the event.The cosy was inspected by the biomed finding debris within the apl bypass valve causing it to stick open intermittently.As the apl bypass valve is open during automatic ventilation this could not be root cause of the reported problem.Based on the available information also a leak at the peep/pmax valve can be excluded as no corresponding error entry was found in the logs.Therefore the most likely root cause for the reported leakage is a fresh gas decoupling valve that temporary remained open maybe also caused by the debris which was found in the cosy.Following the recommendations of the ifu the cosy should be reprocessed on a regular base so that the presence of debris can be avoided.In case of a open fresh gas decoupling valve the fresh gas would flow backwards through the absorber to the ags resulting in a huge leak as reported.Due to the fact that the patient gas/ agent is permanently monitored a corresponding alarm e.G.Low tidal volume and low airway pressure would be issued by the device.That in this case the device did not alarm is only possible if the alarm limits for volume and pressure were not set or if the problem vanished before the fabius alarmed.The device was tested after the case and was returned to use.
|