• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem Injury (2348)
Event Type  Injury  
Manufacturer Narrative
As part of our investigation, olympus made multiple follow ups with the user facility by telephone and in writing in an attempt to gather additional information on the reported event; however, no additional information was obtained.The scope was returned to olympus for evaluation.The evaluation confirmed the reported scope issue.Olympus inspected the scope and found the bending section metal tab sharp and protruding out from the bending section cover near the insertion tube side of the scope.Upon removal of the bending section cover, the bending section skeleton was also found broken.The scope was serviced and returned to the user facility.Based on similar reported events and evaluation findings, the cause of the protruding / lifting skeleton metal tab could be attributed to the operator¿s technique.The original equipment manufacturer (oem) performed investigations related to this device issue.As a result, the oem has conducted a field corrective action including a distribution of ¿instructions for safe use¿ to mitigate the potential risk of patient injury.The ¿instructions for safe use¿ provides several warning statements in an effort to prevent equipment damage and patient injury.¿if any of the following conditions occur during an examination, immediately stop the examination and withdraw the endoscope from the patient.If the up/down angulation control lever does not move.If the angulation control mechanism is not functioning properly.Visually inspect the bending section for no metallic parts protruding from the bending section.Visually inspect the bending section for bends, twist, or other irregularities while the bending section remains straight.Visually inspect the bending section for abnormal bending shape, or other irregularities.Continued use of the endoscope under these conditions could result in patient injury, bleeding, and/or perforation.".
 
Event Description
Olympus was informed that during an unspecified event, patient injury occurred.The patient injury is unknown.It was also reported that the scope was found with a broken rib and punctured a hole in the bending section rubber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
408935-512
MDR Report Key7118197
MDR Text Key94873352
Report Number2951238-2017-00768
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Device Catalogue NumberURF-V2
Device Lot NumberN/A
Other Device ID Number04953170343582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2017
Initial Date Manufacturer Received 11/17/2017
Initial Date FDA Received12/14/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1251-2017
Patient Sequence Number1
-
-