Catalog Number AERO815
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during a tracheobronchial stent placement within the patient's right main bronchus the delivery catheter tip detached.The stent successfully deployed as desired by the physician.The physician retrieved the catheter tip using a scope and graspers.No additional consequences to the patient have been reported.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The device was examined visually.The complaint is confirmed.The root cause is attributed to the manufacturing process.A review of the complaint database was performed and no similar complaints for this lot number were found.A review of the device history record was performed and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEROMINI TRACHEOBRONCHIAL STENT
Type of Device
TRACHEOBRONCHIAL STENT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston TX 77047
Manufacturer Contact
casey hughes
1600 w merit parkway
south jordan, UT 84095
8013164932
MDR Report Key7118269
MDR Text Key95767610
Report Number3010665433-2017-00079
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/31/2020
Device Catalogue NumberAERO815
Device Lot NumberIPX2742U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2017
Initial Date FDA Received12/14/2017
Supplement Dates Manufacturer Received01/29/2018
Supplement Dates FDA Received02/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-