• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LENS INTEGRATED SYSTEM WIFI VERSION; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LENS INTEGRATED SYSTEM WIFI VERSION; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 72204354
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the device kept shorting out during the surgery.A backup device was available to complete the procedure following a short delay.No patient impact reported.
 
Manufacturer Narrative
A visual inspection was performed on the exterior of the product and no physical damage was observed.The ccu failed functional testing with no hd-sdi video output.All other video outputs normal.Cause of dead hd-sdi output is a defective electronic component on the dsp module pcb.Ccu passes functional testing with a known good dsp module pcb installed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENS INTEGRATED SYSTEM WIFI VERSION
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7118522
MDR Text Key94995596
Report Number1643264-2017-01825
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72204354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/01/2017
Initial Date FDA Received12/14/2017
Supplement Dates Manufacturer Received01/23/2018
Supplement Dates FDA Received01/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-