MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Break (1069); High impedance (1291); Unintended Collision (1429); Device Or Device Fragments Location Unknown (2590)
Patient Problem Therapeutic Response, Decreased (2271)
Event Type  Injury  
Manufacturer Narrative
The main component of the system and other applicable components are: product id: 3889-28, lot# va0ge5e, implanted: (b)(6) 2014, explanted: (b)(6) 2017, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacturer¿s representative regarding a patient who was implanted with a neurostimulator for urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor.It was reported that the patient had a loss of efficacy and reported having fallen.The patient had their impedances checked in office and their impedance reading were over 4000 ohms.The physician replaced the lead and the neurostimulator.It was noted that the patient¿s lead was fractured, and it was related to their reported fall.The patient had their lead partially explanted as the lead was fractured from the sacral site after the extended dissection.It was noted that the patient was alive with no injury, and the issue was resolved at the time of the report.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7120819
MDR Text Key94955671
Report Number3004209178-2017-26045
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2015
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/13/2017
Initial Date FDA Received12/15/2017
Date Device Manufactured02/24/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-