MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3058 |
Device Problems
Break (1069); High impedance (1291); Unintended Collision (1429); Device Or Device Fragments Location Unknown (2590)
|
Patient Problem
Therapeutic Response, Decreased (2271)
|
Event Type
Injury
|
Manufacturer Narrative
|
The main component of the system and other applicable components are: product id: 3889-28, lot# va0ge5e, implanted: (b)(6) 2014, explanted: (b)(6) 2017, product type: lead.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer via a manufacturer¿s representative regarding a patient who was implanted with a neurostimulator for urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor.It was reported that the patient had a loss of efficacy and reported having fallen.The patient had their impedances checked in office and their impedance reading were over 4000 ohms.The physician replaced the lead and the neurostimulator.It was noted that the patient¿s lead was fractured, and it was related to their reported fall.The patient had their lead partially explanted as the lead was fractured from the sacral site after the extended dissection.It was noted that the patient was alive with no injury, and the issue was resolved at the time of the report.No further complications were reported.
|
|