Model Number G407211
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 11/20/2017
Event Type  Injury  
Manufacturer Narrative
The reported cardiac tamponade could not be confirmed.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Event Description
Related manufacturer report number: 9680001-2017-00103, 3008452825-2017-00328, 2030404-2017-00056.During an atrial tachycardia procedure, a cardiac tamponade occurred.After isolating the pulmonary veins with the ablation catheter the patient became hypotensive and an echocardiogram revealed a cardiac tamponade for which a pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK¿ XS TRANSSEPTAL NEEDLE, 98 CM LENGTH
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL,
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7122080
MDR Text Key94993537
Report Number3008452825-2017-00327
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2020
Device Model NumberG407211
Device Lot Number6131068
Other Device ID Number05414734205153
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2017
Initial Date FDA Received12/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AGILIS NXT; INQUIRY AFOCUS II; TACTICATH QUARTZ
Patient Outcome(s) Required Intervention;
-
-