MEDTRONIC NAVIGATION, INC. (LOUISVILLE) 5.5/6.0 SOLERA DRIVER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735023
Device Problems Device Operates Differently Than Expected (2913); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
Patient weight not available from the site.No parts have been received by the manufacturer for evaluation.This device is included in the medical device field correction notification, "potential for instrument breaking ¿ medtronic navigated solera screwdrivers" ( (b)(6) 2015).The revised instructions for use (ifu) were also provided with the notification.Part not returned for analysis.
 
Event Description
A medtronic representative reported that, while in a t10-ilium due to degenerative disc disease procedure, the tip of the screwdriver being used separated from the driver.It was reported that the issue occurred when the surgeon was making final adjustments to the large screw for placement.No fragments of the device remained inside of the patient.There was no impact on patient outcome.There was no reported delay to the procedure due to this issue.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5/6.0 SOLERA DRIVER
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7123364
MDR Text Key95073170
Report Number1723170-2017-05146
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Notification
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735023
Device Lot Number131024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/22/2017
Initial Date FDA Received12/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2823/2824-2015
Patient Sequence Number1
Patient Age76 YR
-
-