Model Number N/A |
Device Problem
Component Missing (2306)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Cmp-(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the provisional shim was missing ball bearings.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned parts show signs of repeated use and missing ball bearing.Dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required.The root cause of the reported issue is attributed to design deficiency.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|