Model Number 7209807-
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during testing, the unit's plug end was broken except for one piece.No patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The complaint investigation has concluded that this unit has succumbed to damage to cord connector.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The root cause of the observed condition was determined to be a result of a misuse of the product.The instructions for use (ifu) recommends "careful inspection of the operative headpiece before and after the procedure for possible signs of damage.Immediate detection and repair of minor damage will extend the life of the operative handpiece.¿ damage may occur if the headpiece is handled roughly during use.Excessive forces may result in traumatic insertion, chipping particles of the optical glass, or other damage to the unit.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN NORTH HAVEN-MFG
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN NORTH HAVEN-MFG
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
sharon murphy
60 middletown ave
north haven, CT 06473
2034925267
MDR Report Key7127145
MDR Text Key95240952
Report Number1219930-2017-10149
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209807-
Device Catalogue Number7209807-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received12/18/2017
Supplement Dates Manufacturer Received01/24/2018
Supplement Dates FDA Received02/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-