CYBERONICS, INC VNS THERAPY SYSTEM; STIMULATOR, AUTONOMIC NERVE, IMPLANTED FOR EPILEPSY Back to Search Results
Model Number 106
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Code Available (3191)
Event Date 11/29/2017
Event Type  malfunction  
Event Description
Per report, the vagal nerve stimulator was not functioning properly.A prior evaluation by a cyberonics representative showed impedance values in his system greater than 10,000 ohms-indicating a break in his electrode.So the patient was taken to the or for removal and replacement of the nerve stimulator and lead.The device will be returned to the cyberonics representative.Manufacturer response for cyberonics vagal nerve stimulator, cyberonics vagal nerve stimulator (per site reporter).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VNS THERAPY SYSTEM
Type of Device
STIMULATOR, AUTONOMIC NERVE, IMPLANTED FOR EPILEPSY
Manufacturer (Section D)
CYBERONICS, INC
100 cyberonics blvd
houston TX 77058
MDR Report Key7128024
MDR Text Key95193619
Report Number7128024
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number106
Other Device ID NumberLEAD 01227
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/12/2017
Event Location Hospital
Date Report to Manufacturer12/12/2017
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age51 YR
-
-