• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK ATLAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK ATLAS Back to Search Results
Catalog Number 10309477
Device Problem False Negative Result (1225)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reported that there was some dripping coming from the pipette tip although the pipette tip was not damaged.Siemens service went onsite and the issue was resolved by tightening the tubing to the pipette.The qc was in the expected range and the customer has had no further discrepancies and is operational.
 
Event Description
The customer reported false negative for urine blood on the clinitek atlas vs.2+ on the clinitek advantus and positive on the microscopic exam of the sediment.There was no reported injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK ATLAS
Type of Device
CLINITEK ATLAS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SPARTAN MEDICAL
22740 lunn road
cleveland OH 44149
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key7128997
MDR Text Key96007106
Report Number3002637618-2017-00153
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K932674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10309477
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/08/2017
Initial Date FDA Received12/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-