Model Number 60-03-75
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump console.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative confirmed the reported issue and was unable to correct the problem by rebooting the device.The service representative re-seated all connections on the back plane, cpu and interface boards, but the issue persisted.The cpu board was replaced and the issue was resolved.Subsequent testing found the device to be functioning according to specifications.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
 
Event Description
Livanova (b)(4) received a report that the display of the centrifugal pump console became blank and the background turned blue during setup.The device was restarted and began to function as expected.There was no patient involvement.
 
Manufacturer Narrative
The defective board was returned to livanova (b)(4) for further investigation.During the investigation the reported failure could be confirmed and traced back to a defective memory on the board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP CONSOLE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
144001 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7135926
MDR Text Key95751889
Report Number9611109-2017-01034
Device Sequence Number1
Product Code KFM
UDI-Device Identifier04033817901020
UDI-Public010403381790102011170614
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-03-75
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2017
Initial Date FDA Received12/20/2017
Supplement Dates Manufacturer Received02/14/2018
Supplement Dates FDA Received04/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-